Contamination of circuit boards can bring about severe degradation of insulation resistance and dielectric strength. Cleanliness of completed circuit boards is, therefore, of vital interest.

For those companies who have established circuit board cleaning procedures, the MIL Spec P-28809 has been used as a guideline for control. Now a simple "on line" test for the relative measurement of ionic contamination has been developed.

This fast and economical method for testing circuit board cleanliness uses an Ultrameter II™ 4P or 6P, a suitable container, and a mixture of Dl (deionized) water and alcohol. The procedure is as follows:

   1. Mix a stock quantity of solution using 25 parts by volume of Dl water and 75 parts by volume of 99% isopropyl alcohol. The conductivity, measured with the Ultrameter II 4P or 6P should be a maximum of 0.166 micromhos/microseimens/cm.

   2. Measure out an amount of the water/alcohol mixture equal to 100 ml per 10 square inches of circuit board surface to be tested (considering both sides of the board but not components), and add 60 ml additional. In other words: 2(L X W) (10 ml) + 60 ml = total solution needed.

   3. Fill a poly "zip-lock" bag or other suitable plastic or glass container with the measured water/alcohol solution.

   4. Using the measured water/alcohol solution in the poly bag, rinse out the Ultrameter II's cell cup three (3) times, discarding the rinse solution each time. Fill the instrument cell cup a fourth time and take a meter reading. This value should be 0.166 micromhos/microseimens/cm or less and is the very clean control (or "comparison") reading for the test.

   5. Being very careful not to contaminate the PCB, totally immerse the circuit board in the solution. Seal bag. Allow it to soak for three (3) minutes with mild agitation.

   6. At the conclusion of the soaking, pour the solution directly into the instruments cell cup four (4) times; take the fourth reading.

   7. Compare the control reading in Step 4 with the reading taken in Step 6 (The higher the difference between the two readings, the greater the ionic contamination). Record this final extract reading for comparison with other boards tested in the same manner.

The level of cleanliness needed or desired can only be determined by each individual company.

 

Mil Spec P-28809 can be used as a guideline, or standards can be established based upon available data. In either event, the comparative method using the Myron L Company's Ultrameter II will assist in the determination of that level of cleanliness.



1-866-907-5459

Water Industry News
Worldwide Shipping

Get the new Myron L Meters mobile app.
 
Android app on Google Play
Ask An Expert

Enter Your Email

Get special offers and discounts on items you already purchase anyway...

Connect With Us Online

Share insights with other water treatment professionals in various industries, access to industry updates, and more!

Connect with us on Facebook Connect with us on LinkedIn Connect with us on Twitter Watch us on YouTube Pin us on Pinterest

How Do You Like Our Website?

We'd like to get your feedback. We are dedicated to exceptional service, so tell us what you think! Tell Us Now

  • We're ShopWiki Approved!
  • MyronLMeters.com is UpFront